Product Description
The Parallel/Depth Guide Short is a precision implantology instrument designed for accurate depth measurement and parallel alignment verification during implant osteotomy preparation and dental implant surgical procedures.
The instrument is intended for use during the initial preparation of the implant bone bed to assist clinicians in evaluating drilling depth, implant trajectory, and parallel positioning in relation to adjacent teeth, implants, anatomical structures, and the mandibular nerve.
The guide features a precision working section with a length of 10 mm and threaded reference markings with a 1 mm pitch, allowing highly accurate radiographic depth evaluation and positional assessment during surgical procedures.
After preparation of the initial osteotomy with a pilot drill, the Parallel/Depth Guide Short is inserted into the prepared implant site to enable radiographic examination using periapical or panoramic imaging techniques. This allows clinicians to verify implant positioning, evaluate proximity to anatomical structures, and confirm parallel drilling orientation before continuing osteotomy preparation.
Manufactured from high-quality medical-grade titanium alloy Ti6Al4V ASTM F136, the instrument provides excellent dimensional precision, corrosion resistance, biocompatibility, mechanical durability, and long-term clinical reliability. The smooth polished surface supports effective cleaning and sterilization procedures.
The Parallel/Depth Guide Short is intended for use in oral implantology procedures requiring controlled implant positioning and accurate radiographic depth verification.
Suitable for:
- Implant depth verification
- Parallel drilling evaluation
- Implant osteotomy preparation
- Radiographic implant positioning
- Implant surgical planning
- Oral implantology procedures
The precision threaded design allows accurate radiographic measurement and controlled evaluation of implant positioning during surgical procedures.
Specifications
- Instrument type: Parallel/depth guide short
- Working length: 10 mm
- Design: Threaded radiographic measuring guide
- Thread pitch: 1 mm
- Material: Medical-grade titanium alloy Ti6Al4V ASTM F136
- Surface finish: Smooth polished surface
- Application: Implant depth and parallel verification
- Sterilization: Fully autoclavable
Indications for Use
Intended for:
- Implant osteotomy depth measurement
- Parallel drilling verification
- Radiographic implant positioning evaluation
- Implant surgical planning
- Bone bed preparation procedures
- Oral implantology applications
Sterilization
The instrument must be sterilized before use in an autoclave at: 135°C (274°F)
Advantages
- Accurate implant depth verification
- Precise parallel drilling evaluation
- 1 mm threaded reference markings
- Improved radiographic visualization
- Supports controlled implant positioning
- High-strength titanium alloy construction
- Excellent biocompatibility and corrosion resistance
- Smooth easy-to-clean surface
- Reliable clinical performance
- Fully autoclavable
- Designed for predictable implant surgical procedures
Reviews
There are no reviews for this product, be the first to leave your review.
Answers & questions
No questions about this product, be the first and ask your question.
Parallel/Depth Guide Short
Product Description
The Parallel/Depth Guide Short is a reusable implantology instrument intended for accurate depth measurement and parallel alignment verification during implant osteotomy preparation and dental implant surgical procedures.
The instrument is designed for insertion into the prepared implant osteotomy after pilot drilling in order to assist clinicians in evaluating drilling depth, implant trajectory, and parallel positioning relative to adjacent teeth, implants, and anatomical structures.
The working section has a length of 10 mm and features threaded reference markings with a 1 mm pitch, allowing precise radiographic evaluation and depth measurement during surgical procedures.
The instrument is intended for use together with radiographic imaging techniques, including periapical and panoramic radiographs, to verify implant positioning before continuation of osteotomy preparation.
Manufactured from medical-grade titanium alloy Ti6Al4V ASTM F136, the instrument provides excellent dimensional precision, corrosion resistance, biocompatibility, mechanical durability, and long-term clinical reliability.
Intended Use
The Parallel/Depth Guide Short is intended for:
- Implant osteotomy depth verification
- Parallel drilling evaluation
- Radiographic implant positioning assessment
- Implant surgical planning
- Bone bed preparation procedures
- Oral implantology applications
Indications for Use
The instrument is indicated for use in dental implant surgical procedures requiring controlled verification of osteotomy depth, implant trajectory, and parallel positioning before final implant site preparation.
Contraindications
Do not use the instrument:
- For procedures other than intended dental implant applications
- If the instrument is damaged, bent, worn, or corroded
- Without proper sterilization before clinical use
- In cases where radiographic evaluation is contraindicated
Warnings
- The instrument must be used only by qualified dental professionals trained in implantology procedures.
- Improper interpretation of radiographic measurements may compromise implant positioning.
- Inspect the instrument before each use for wear, deformation, or damage.
- Do not apply excessive force during insertion into the osteotomy site.
- Use only according to accepted implant surgical protocols.
Precautions
- Sterilize the instrument before first use and before each subsequent use.
- Use appropriate radiographic techniques for accurate evaluation.
- Ensure the osteotomy site is properly prepared before inserting the guide.
- Clean the instrument immediately after use to prevent residue accumulation.
- Follow standard implantology and oral surgical procedures.
Directions for Use
1. Clean and sterilize the instrument before use.
2. Prepare the initial osteotomy using the pilot drill according to the implant surgical protocol.
3. Insert the Parallel/Depth Guide Short into the prepared osteotomy site.
4. Perform radiographic imaging using periapical or panoramic radiographs.
5. Evaluate osteotomy depth, implant positioning, and drilling parallelism using the 1 mm threaded reference markings.
6. Confirm safe distance from adjacent anatomical structures before continuing osteotomy preparation.
7. Remove the guide carefully after radiographic evaluation.
8. Continue implant site preparation according to the surgical protocol.
9. After completion of the procedure, clean, disinfect, and sterilize the instrument before reuse.
Cleaning and Disinfection
Immediately after use:
- Remove visible debris and biological contaminants.
- Rinse thoroughly under running water.
- Clean manually or using an ultrasonic cleaner with approved medical instrument detergent.
- Dry the instrument completely before sterilization.
Sterilization
- The instrument must be sterilized before use in an autoclave at: 135°C (274°F)
- Sterilization must be performed according to standard dental sterilization procedures and local regulations.
Storage
- Store in a clean, dry environment.
- Protect from moisture and contamination.
- Store in appropriate sterilization packaging until use.