Product Description
Implant Drive Key with Hexagon 2.1
The Implant Drive Key with Hexagon 2.1 is a precision surgical instrument designed for manual placement and removal of narrow dental implants with Internal Hex H2.1 and Conical Connection CH2.1 interfaces. The instrument provides secure implant engagement, accurate torque transfer, and reliable control during implant surgery procedures.
The lower working part of the key has a length of 18 mm and features a precision 2.1 mm hexagonal connection engineered for stable insertion into compatible implant systems with:
- Internal Hex H2.1 connection
- Conical Connection CH2.1
The key may also be inserted into the implant carrier from the end side for installation of the implant into the prepared bone bed together with the implant carrier.
The upper part of the instrument incorporates a universal 6.35 mm hexagonal connection compatible with:
- Hand drive wrenches
- Ratchet wrenches (446301, etc.)
- Hollow wrenches (446303, etc.)
Manufactured from high-quality 420 series medical stainless steel, the Implant Drive Key provides excellent mechanical strength, corrosion resistance, dimensional precision, and long-term clinical durability. The smooth outer surface improves handling and simplifies cleaning and sterilization procedures.
The instrument features elevated orientation edges that assist clinicians in visually controlling implant hexagon positioning during placement procedures.
For optimal prosthetic orientation, it is recommended to position the flat side of the facet toward the vestibular side, allowing proper use of standard prosthetic abutments.
The key also includes two circular depth markings indicating implant immersion depth relative to the bone and gingival level:
- First mark: 3 mm
- Second mark: 5 mm
Suitable for:
- Manual implant placement
- Implant removal procedures
- Narrow implant surgery
- Implant positioning and orientation
- Implant restorative preparation
- Oral implantology procedures
Specifications
- Instrument type: Implant drive key
- Hex connection: 2.1 mm
- Working length: 18 mm
- Upper connection: 6.35 mm hex
- Material: 420 series medical stainless steel
- Surface finish: Smooth polished surface
- Sterilization: Fully autoclavable
Compatible Implant Systems
- Internal Hex H2.1 implants
- Conical Connection CH2.1 implants
Indications for Use
Intended for manual installation and removal of compatible implants during:
- Implant surgery
- Implant positioning
- Implant insertion procedures
- Oral surgical implant procedures
Sterilization
The instrument must be sterilized before use in an autoclave at: 135°C (274°F)
Advantages
- Precision 2.1 mm implant connection
- Compatible with H2.1 and CH2.1 implant systems
- Universal 6.35 mm ratchet compatibility
- Depth control reference markings
- Visual implant orientation guidance
- High-strength 420 medical stainless steel construction
- Excellent corrosion resistance
- Smooth ergonomic surface design
- Reliable torque transfer and implant control
- Fully autoclavable
- Designed for predictable implant surgical procedures
Reviews
There are no reviews for this product, be the first to leave your review.
Answers & questions
No questions about this product, be the first and ask your question.
General Information
Before using DMi products, the dentist, maxillofacial surgeon, or dental technician should carefully read all indications, contraindications, recommendations, warnings, and instructions, as well as all additional product documentation including surgical manuals, implant catalogs, and technical specifications.
DMi recommends professional education and training in implantology before clinical use.
Detailed product information is available through authorized DMi representatives and on the company website:
www.dmimplant.online
Product Description
The Implant Drive Key with Hexagon 2.1 is a precision surgical instrument designed for manual placement and removal of narrow implants with:
- Internal Hex H2.1 connection
- Conical Connection CH2.1
- The lower working part of the key has a length of: 18 mm
The lower part of the instrument features a precision 2.1 mm hexagonal connection designed for insertion into compatible implant systems.
The upper part of the key contains a universal: 6.35 mm hexagonal connection
The instrument is compatible with:
- Hand drive wrenches
- Ratchet wrenches (446301, etc.)
- Hollow wrenches (446303, etc.)
The key may also be inserted into the implant carrier from the end side for installation of the implant into the prepared bone bed together with the implant carrier.
The instrument is manufactured from high-quality 420 series medical stainless steel and features a smooth polished outer surface.
The key incorporates:
- Elevated orientation edges for visual positioning of the implant hexagon
- Two circular depth markings: 3 mm depth mark and 5 mm depth mark
For proper prosthetic orientation, it is recommended to position the flat side of the facet toward the vestibular side to facilitate use of standard prosthetic abutments.
Intended Use
The Implant Drive Key with Hexagon 2.1 is intended for:
- Manual implant placement
- Implant removal
- Implant positioning and orientation
- Implant surgical procedures
- Implant carrier-assisted insertion procedures
Indications
The instrument is indicated for:
- Placement of narrow implants
- Removal of compatible implants
- Surgical implant positioning
- Implant orientation during surgery
- Implant insertion with implant carriers
Compatible with:
- Internal Hex H2.1 implants
- Conical Connection CH2.1 implants
Contraindications
Do not use in cases of:
- Damaged or worn instrument connections
- Use with incompatible implant systems
- Improper sterilization
- Improper implant selection
Warnings
- The product is supplied non-sterile.
- Sterilization before use is mandatory.
- Only qualified dental professionals should use the instrument.
- Excessive torque may damage implants or implant components.
- Inspect the instrument before each use for wear, deformation, or damage.
- Use only compatible implant systems and accessories.
Clinical Procedure
1. Preparation
- Sterilize the Implant Drive Key before use.
- Verify compatibility between the implant system and the instrument.
2. Instrument Connection
Connect the upper 6.35 mm hexagonal part to:
- Hand drive wrench
- Ratchet wrench
- Hollow wrench
- Compatible surgical driver system
3. Implant Engagement
- Insert the 2.1 mm hexagonal working part into the internal implant connection.
- Ensure stable and complete engagement before torque application.
4. Implant Orientation
- Use the elevated orientation edges for visual positioning of the implant hexagon.
- Position the flat side of the facet toward the vestibular side to facilitate use of standard abutments.
5. Implant Placement
- Carefully insert the implant into the prepared osteotomy site.
- Monitor implant depth using the circular depth markings:
- First mark: 3 mm
- Second mark: 5 mm
6. Final Verification
Verify:
- Implant stability
- Implant depth
- Proper prosthetic orientation
- Correct implant positioning
7. Implant Removal (if required)
- Engage the implant connection securely.
- Carefully reverse the implant using controlled manual torque.
Sterilization
Before use, sterilize the instrument in an autoclave at: 135°C (274°F)
Storage
Store in a clean, dry environment away from contamination, moisture, and direct sunlight.