Product Description
The Closed Transfer for Standard Connector Abutment is a precision impression component designed for accurate transfer of the intraoral position of osseointegrated implants and standard connector abutments to the working model during prosthetic procedures.
The transfer is compatible with standard connector abutments (Ref. 604040, etc.) and standard titanium connectors (Ref. 604040, etc.). Designed without an anti-rotation feature, it connects directly to the standard connector using fixation screw 711729, providing stable positioning and reliable impression transfer accuracy.
The component is intended for use with the closed tray impression technique and allows precise reproduction of implant-abutment positioning in laboratory workflows for prosthetic restorations.
Manufactured from medical-grade titanium alloy Ti6Al4V ASTM F136, the transfer offers excellent dimensional stability, biocompatibility, corrosion resistance, and durability during clinical use.
Key Features
- Designed for closed tray impression technique
- Compatible with standard connector abutments
- Non-anti-rotational design
- Direct connection using fixation screw 711729
- Precision-machined for accurate impression transfer
- High dimensional stability and durability
Available Versions / References
- Standard connector abutment: 604040, etc.
- Fixation screw: 711729
Technical Specifications
- Diameter: 4.5 mm
- Height: 10 mm
- Hex connection: 1.25 mm
Material
Manufactured from medical-grade titanium alloy Ti6Al4V ASTM F136, providing:
- High mechanical strength
- Excellent biocompatibility
- Corrosion resistance
- Precision and durability for clinical use
Clinical / Laboratory Application
- Closed tray implant impression procedures
- Transfer of implant-abutment position to laboratory model
- Prosthetic workflows for osseointegrated implants
- Accurate reproduction of implant positioning
Sterilization
Before clinical use, the product must be sterilized in an autoclave at 135°C / 274°F.
Advantages
- Accurate transfer of implant position
- Stable connection to standard connector abutment
- Easy handling with 1.25 mm hex driver
- Durable and biocompatible titanium construction
- Supports accurate and predictable prosthetic outcomes
Reviews
There are no reviews for this product, be the first to leave your review.
Answers & questions
No questions about this product, be the first and ask your question.
Closed Transfer for Standard Connector Abutment
Intended Use
The Closed Transfer is intended for:
- Closed tray implant impression procedures
- Transfer of implant-abutment position to the dental laboratory model
- Prosthetic restoration workflows for osseointegrated implants
Indications
The product is indicated for use with:
- Standard connector abutments (Ref. 604040, etc.)
- Osseointegrated dental implant systems
- Prosthetic impression procedures using closed tray techniques
Contraindications
Do not use the product in:
- Patients with known allergy or hypersensitivity to titanium alloys
- Clinical situations incompatible with closed tray impression procedures
- Cases where the component is damaged, deformed, or shows signs of wear
Warnings
- Product is supplied non-sterile.
- Sterilization before clinical use is mandatory.
- Use only compatible prosthetic components and instruments.
- Improper seating of the transfer may lead to inaccurate impressions and prosthetic misfit.
- Do not use excessive force during tightening.
- Single-patient use only unless local regulations specify otherwise.
Precautions
- Verify compatibility with the implant system before use.
- Inspect the component before clinical application.
- Ensure complete seating of the transfer onto the abutment.
- Use only a 1.25 mm hex driver for tightening procedures.
- Follow standard implant prosthetic protocols and infection control procedures.
Instructions for Clinical Use
1. Preparation
- Remove the healing component or temporary restoration from the implant abutment.
- Clean and dry the prosthetic connection area.
2. Placement of the Closed Transfer
- Position the Closed Transfer onto the standard connector abutment.
- Secure the transfer using fixation screw 711729.
- Tighten using a 1.25 mm hex driver until stable seating is achieved.
3. Impression Procedure
- Verify correct positioning and stability of the transfer.
- Prepare the impression tray for the closed tray technique.
- Inject impression material around the transfer and seat the tray.
- Allow the impression material to fully polymerize according to the manufacturer’s instructions.
4. Removal
- Remove the impression tray from the patient’s mouth.
- Unscrew and remove the transfer from the mouth.
- Attach the laboratory analogue to the transfer.
- Reposition the transfer assembly into the impression for model fabrication.
Cleaning and Sterilization
The product must be cleaned and sterilized before use.
- Sterilization Parameters
- Autoclave sterilization
- Temperature: 135°C / 274°F
Storage
- Store in a clean, dry environment.
- Protect from contamination, moisture, and mechanical damage.
- Avoid exposure to corrosive substances.