Product Description
The Big Aesthetic Multi-Unit M1.6 Set is a comprehensive prosthetic solution designed for advanced implant restorations utilizing Multi-Unit Abutment systems. The kit provides clinicians and dental technicians with a complete assortment of restorative components required for temporary and permanent screw-retained and cement-retained prosthetic rehabilitations.
The system is specifically developed to maximize clinical flexibility, simplify restorative workflows, and provide immediate access to a wide range of prosthetic options during implant treatment. The extensive selection of Multi-Unit components allows efficient management of various clinical situations, reducing treatment delays and improving restorative predictability.
Manufactured from high-quality medical-grade materials, the components provide excellent dimensional accuracy, mechanical strength, corrosion resistance, and long-term clinical reliability.
The Big Aesthetic Multi-Unit M1.6 Set is suitable for single-unit restorations, multi-unit bridges, full-arch rehabilitations, immediate loading protocols, temporary restorations, and definitive prosthetic solutions.
Set Components
1. Straight Aesthetic Multi-Unit with a shoulder height of 0.5mm, 1.5mm, 2.5mm, 3.5mm, and 4.5mm, 10 pieces of each height (50 pieces).
2. Angulated 17° degrees and 30° degrees Aesthetic Multi-Unit with a shoulder height of 1mm, 2mm, 3mm, and 4mm, 2 pieces of each height (14 pieces).
3. Healing Caps for Multi-Unit (20 pieces).
4. Scan Body Transfer for Multi-Unit (9 pieces) .
5. Open Tray Transfers for Multi-Unit.
6. Titanium Screw M1.6 for Multi-Unit (20 pieces) .
7. Titanium Base, 4mm high, for Multi-Unit (15 pieces).
8. Titanium Base, 10mm high, for Multi-Unit (15 pieces).
9. Plastic Sheath for casting for Multi-Unit (10 pieces) .
10. Titanium Sheath for turning for Multi-Unit (10 pieces).
11. Multi-Unit Analog (8 pieces).
12. Multi-Unit Digital Analog (8 pieces).
13. Hex Key for Multi-Unit and Screws.
Suitable for:
- Temporary restorations
- Permanent restorations
- Screw-retained restorations
- Cement-retained restorations
- Full-arch implant restorations
- Implant-supported bridges
- Immediate loading protocols
- CAD/CAM restorations
- Digital implant workflows
- Conventional implant prosthetics
- Oral implantology applications
Specifications
- Product type: Multi-Unit prosthetic kit
- Screw system: M1.6
- Abutment types: Straight and angulated
- Angulations: 17° and 30°
- Restoration types: Screw-retained and cement-retained
- Workflow compatibility: Conventional and digital
- Material: Medical-grade titanium and laboratory materials
- Sterilization: Fully autoclavable components
Indications for Use
Intended for fabrication and fixation of implant-supported restorations during:
- Single-unit restorations
- Multi-unit restorations
- Full-arch rehabilitations
- Immediate loading restorations
- Temporary prosthetic procedures
- Definitive prosthetic restorations
- Digital restorative workflows
Sterilization
All reusable components must be sterilized before use in an autoclave at: 135°C (274°F)
Advantages
- Complete Multi-Unit restorative solution
- Supports both screw-retained and cement-retained restorations
- Wide selection of straight and angulated abutments
- Multiple gingival height options
- Supports digital and conventional workflows
- Includes components for temporary and permanent restorations
- Comprehensive laboratory and clinical prosthetic kit
- Ideal for full-arch and multi-unit restorations
- M1.6 prosthetic screw system
- Compatible with CAD/CAM workflows
- High-strength medical-grade materials
- Excellent corrosion resistance
- Organized restorative workflow
- Reduces treatment time and inventory requirements
- Designed for predictable implant prosthetic outcomes
Reviews
There are no reviews for this product, be the first to leave your review.
Answers & questions
No questions about this product, be the first and ask your question.
Big Aesthetic Multi-Unit M1.6 Set
Intended Use
The Big Aesthetic Multi-Unit M1.6 Set is intended for prosthetic rehabilitation of dental implants using Multi-Unit Abutment systems. The set provides a comprehensive selection of components for fabrication of temporary and permanent screw-retained and cement-retained restorations.
The kit is designed for use by dentists and dental technicians during clinical and laboratory procedures involving implant-supported restorations, including conventional and digital workflows.
Indications
The Big Aesthetic Multi-Unit M1.6 Set is indicated for:
- Single-tooth restorations
- Multiple-unit restorations
- Implant-supported bridges
- Full-arch restorations
- Screw-retained restorations
- Cement-retained restorations
- Immediate loading restorations
- Temporary restorations
- Permanent restorations
- CAD/CAM restorations
- Digital implant workflows
Clinical Procedure
1. Multi-Unit Abutment Selection
- Select the appropriate straight or angulated Multi-Unit Abutment based on:
- Implant position
- Soft tissue height
- Prosthetic emergence profile
Implant angulation
Verify complete seating of the abutment on the implant platform.
Tighten according to the implant system protocol.
2. Soft Tissue Healing
When soft tissue healing is required:
- Select the appropriate Healing Cap.
- Install the Healing Cap onto the Multi-Unit Abutment.
- Tighten using the supplied hex key.
- Allow soft tissue maturation according to the treatment plan.
3. Impression Procedure
Open Tray Technique
- Install the Open Tray Transfer onto the Multi-Unit Abutment.
- Verify complete seating.
- Take the implant-level impression according to standard prosthetic protocols.
- Remove the impression with the transfer attached.
Digital Impression Technique
- Install the Scan Body Transfer onto the Multi-Unit Abutment.
- Verify complete seating.
- Perform intraoral scanning.
- Export digital data for CAD/CAM restoration design.
4. Laboratory Procedures
Model Fabrication
- Attach the Multi-Unit Analog or Digital Analog to the impression transfer or digital model.
- Fabricate the working model according to laboratory protocol.
Restoration Fabrication
The following components may be used:
- Titanium Base 4 mm
- Titanium Base 10 mm
- Plastic Sheath for Casting
- Titanium Sheath for Milling
Fabricate the restoration according to conventional or CAD/CAM laboratory procedures.
5. Temporary Restoration
- Fabricate the temporary restoration using the selected prosthetic component.
- Install onto the Multi-Unit Abutment.
- Secure using the M1.6 Titanium Screw.
- Verify fit and occlusion.
6. Permanent Restoration
Verify passive fit of the definitive restoration.
Seat the restoration onto the Multi-Unit Abutments.
Insert the M1.6 Titanium Screw.
Tighten according to the recommended prosthetic protocol.
Verify:
Stability
Passive fit
Occlusion
Esthetics
Seal the screw access channel according to standard restorative procedures.
Removal Procedure
Remove restorative material from the screw access channel.
Loosen the M1.6 Titanium Screw using the supplied hex key.
Remove the restoration.
Inspect all components before reuse.
Cleaning and Sterilization
Remove all biological and laboratory contaminants.
Clean using approved medical device cleaning procedures.
Rinse thoroughly with distilled or deionized water.
Dry completely before sterilization.
Sterilize all reusable components before use in a steam autoclave: Temperature: 135°C (274°F)
Follow validated sterilization procedures and local regulatory requirements.
Warnings and Precautions
For professional dental use only.
Use only with compatible Multi-Unit implant systems.
Verify complete seating of all components before tightening.
Follow recommended torque values for prosthetic screws and abutments.
Do not use damaged or deformed components.
Components showing wear or damage should be replaced.
Ensure passive fit of all restorations before final fixation.
Storage
Store components in a clean, dry environment protected from contamination, moisture, and mechanical damage.