Product Description
Titanium Abutment for Short Implant H2.43
The Titanium Abutment for Short Implant H2.43 is a precision-engineered prosthetic component designed specifically for cement-retained restorations on short dental implants. Manufactured from medical-grade Ti6Al4V ASTM F136 titanium alloy, this abutment provides exceptional strength, biocompatibility, and long-term clinical reliability.
Featuring a straight profile with a 2.43 mm anti-rotational hex connection, the abutment ensures accurate positioning and a secure implant-abutment interface. Its design is optimized for use with short implant systems, providing a stable prosthetic foundation while supporting predictable restorative outcomes.
The abutment connects directly to compatible H2.43 short implants using fixation screw 701207. It is intended for cement-retained restorations in single-unit and multiple-unit treatment applications and is compatible with a range of short implant diameters and lengths.
Advantages
- Manufactured from medical-grade Ti6Al4V ASTM F136 titanium alloy.
- Designed specifically for short implant restorations.
- Anti-rotational 2.43 mm hex connection for accurate positioning and stability.
- Suitable for cement-retained implant-supported restorations.
- Excellent mechanical strength and corrosion resistance.
- Bio-compatible material suitable for long-term clinical use.
- Direct connection to H2.43 short implant platform.
- Compatible with multiple implant diameters and short implant lengths.
- Provides a stable prosthetic platform for predictable restorative outcomes.
- Supports both single-unit and multiple-unit restorations.
Specifications
- Connection: H2.43 Internal Hex 2.43 mm
- Diameter: 4.5 mm
- Height: 9 mm
- Material: Ti6Al4V ASTM F136 Titanium Alloy
- Fixation Screw: 701207
Compatible Implants
- UNIQUE Short Implants with Lengths: 5 mm, 6 mm, 7 mm and Diameters: Ø3.75 mm, Ø4.2 mm, Ø5.0 mm, Ø6.0 mm
- PRIMA Short Implants with Lengths: 5 mm, 6 mm, 7 mm and Diameters: Ø3.75 mm, Ø4.2 mm, Ø5.0 mm, Ø6.0 mm
Indications
- Cement-retained implant-supported restorations.
- Restorations on short implant systems.
- Single-unit and multiple-unit prosthetic restorations.
- Implant prosthetic rehabilitation where reduced implant length is indicated.
Sterilization
The product is supplied non-sterile. Prior to clinical use, sterilization in an autoclave at 135°C (274°F) is required according to validated sterilization procedures.
Reviews
There are no reviews for this product, be the first to leave your review.
Titanium Abutment for Short Implant H2.43
Intended Use
The Titanium Abutment for Short Implant H2.43 is intended for use with compatible short dental implants to support cement-retained implant-supported restorations.
Contraindications
Do not use:
- With implant systems not listed in these instructions.
- If the component is damaged, deformed, or contaminated.
- In patients with conditions that contraindicate implant treatment.
- When compatibility with the implant system cannot be verified.
- In cases of insufficient implant stability or failed osseointegration.
Warnings
- This product is intended for use only by qualified dental professionals.
- Use only original compatible components and fixation screws.
- Failure to properly seat the abutment may result in mechanical complications, component loosening, or restoration failure.
- Improper prosthetic procedures may compromise the fit, function, and longevity of the restoration.
- The product is intended for single-patient use only.
Precautions
- Verify implant compatibility before use.
- Inspect the abutment and fixation screw prior to use.
- Confirm complete seating of the abutment before tightening the fixation screw.
- Verify proper fit clinically and radiographically.
- Follow accepted clinical protocols for implant restoration and maintenance.
- Avoid excessive occlusal loading, particularly in restorations supported by short implants.
- Maintain traceability of the product lot number in patient records.
Instructions for Use
Clinical Procedure
1. Verify compatibility of the abutment with the selected short implant system.
2. Remove the healing abutment or cover screw and clean the implant connection.
3. Inspect the internal implant hex and ensure it is free of debris and contamination.
4. Seat the Titanium Abutment for Short Implant H2.43 into the implant connection and verify complete seating.
5. Secure the abutment using fixation screw 701207 and tighten according to the implant system protocol.
6. Verify the position and fit of the abutment clinically and radiographically.
7. Prepare the abutment as required for the planned restoration.
8. Fabricate or place the definitive restoration.
9. Verify occlusion, stability, and soft tissue adaptation.
10. For cement-retained restorations, remove all excess cement.
11. Perform routine follow-up examinations according to accepted clinical protocols.
Sterilization
- The product is supplied non-sterile.
- Before clinical use, clean and sterilize the abutment using a validated steam sterilization cycle. Autoclave sterilization at 135°C (274°F) is required prior to placement.
- Do not use the product if sterility cannot be assured.
Storage
Store in a clean, dry environment in the original packaging. Protect from contamination, excessive heat, direct sunlight, moisture, and mechanical damage.